A SUBMERSIBLE PROBE FOR MEASURING THE AMOUNT OF DISSOLVED OXYGEN IN WATER

O2 IFF PROBE XB4-S O2

Effective and easy measurement at fish farms

This probe consists of a stainless-steel head with the option of a replaceable sensor (CAP), a body with integrated electronics and a durable cable. Optical oxygen probes are distinguished by their extremely long service-life with minimum calibration requirements. The silicone membrane means that XB4-S probes also have much higher reaction-capabilities than competitive probes with Teflon elements.

Download datasheet Start your automation
XB4-S-O2
SIMPLE CONTROL

Watch video about
probe calibration

The calibration and setup processes for your probes are intended to be as straightforward as possible. You will instantly understand how to correctly configure your probes so that they will precisely match the conditions you will need to have for the greatest aquaculture environment! You will be able to handle it on your own after watching this brief and simple video.

Advantages

The XB4-S probe is crucial from the aspect of energy savings linked to operation of the bellows and aeration of the tanks. This type of measurement ensures the optimum concentration of dissolved oxygen in the water depending on the species of fish being farmed.

  • 01 Highly-durable sheath intended for fresh water and salt water application
  • 02 Measurements taken independently of water currents around the sensor
  • 03 Robust and easily removed sensor – CAP
  • 04 Minimum operating costs
  • 05 Not sensitive to the presence of H2S or CO2 gas

Structure

XB4-S-O2
Case study

IFF installation at
a Lobster farm, Selsey UK

We achieved several firsts thanks to the installation of the IFF System in Selsey. Not only was it the first commercial installation in the UK, but it was also our first collaborative project with Oxysystems Ltd.

Read full case study
Basic technical parameters
Power supply external source 18 ÷ 35V DC/ 30 mA
Signal output 4 ÷ 20 mA + MODBUS
Oxygen concentration measuring range 0,0 ÷ 200,0 %, 0,00 ÷ 20,00 mg/l
Oxygen concentration measuring accuracy 1 % of range; ± 1dig.
Response Time T98% < 60 s
Temperature compensation automatic 0 ÷ 50 °C
Temperature measuring range 0 ÷ 50 °C
Temperature measurement accuracy ± 0,2 °C; ± 1dig.
Temperature sensor Ni 1000
Measurement principle oxygen dependent luminescence quenching
Protection rating IP 68
Pressure range 0 to 3,5 bar (0 ÷ 35 m water column)
Operating temperature range 0 ÷ 50 °C
Storage temperature range -20 ÷ 70 °C
Dimensions (probe) 270 x 36 mm
Dimensions (box) 320 x 200 x 75 mm
Material stainless steel 1,4404, PVC-U black, PUR (cable)
Weight (probe + 10m cable) 1,7 kg
Oxygen consumption none
Required flow none
Maximum immersion depth 30 m